GUI reference
R&S
®
ZNA
622
User Manual 1178.6462.02 ─ 20
Background information
Refer to
Chapter 4.7.26, "Direct generator/receiver access"
on page 277 and
ter 4.7.29, "Direct IF access"
Receiver Input
Determines how the MEAS and REF signals are routed to the respective receivers.
Depending on the selected path, the hardware and factory correction values are set. If
"Port" is used, also the factory calibration is applied. This setting ensures reasonably
corrected data with the different cabling options.
"Port"
Via the standard port and with standard jumpering of the
connectors.
"Direct
Access"
Via the direct generator/receiver access connectors (see
ter 4.7.26, "Direct generator/receiver access"
"Reverse Cou-
pler"
Via the standard port but with reversed coupler configuration (see
"Reverse Coupler Configuration"
"Source Moni-
tor"
Via the direct source monitor access connector (see
"Direct source monitor access"
Can only be selected for analyzer ports 1 and 3, and only if the
optional source monitor access is available for this port.
"Rear IN"
Via the direct IF access connectors (see
For "Direct Access", "Reverse Coupler", and "Source Monitor" access, the LED above
the direct generator/receiver access connectors is switched on
For analyzer ports that are used as converter ports, the "Receiver Input" is configured
in the
Converter Configuration dialog
.
Remote command:
[SENSe<Ch>:]PATH<Pt>:DIRectaccess
SOURce<Ch>:PATH<Pt>:DIRectaccess
[SENSe<Ch>:]PATH<Pt>:MEASurement:DIRectaccess?
[SENSe<Ch>:]PATH<Pt>:REFerence:DIRectaccess?
[SENSe<Ch>:]PATH<Pt>:MEASurement:IFSWitch?
Rear Out
If checked, the
connectors on the rear panel are used as (additional)
output.
Remote command:
[SENSe<Ch>:]PATH<Pt>:IFSWitch
[SENSe<Ch>:]PATH<Pt>:MEASurement:IFSWitch?
[SENSe<Ch>:]PATH<Pt>:REFerence:IFSWitch?
Auto Determine IF Out
This checkbox is only enabled, if
is used.
Channel Config softtool