GUI reference
R&S
®
ZNA
548
User Manual 1178.6462.02 ─ 20
Start Cal Sweep
Start the calibration sweep for the selected port and power calibration settings and
close the dialog. The calibration is performed as described in
on page 194. No additional calibration settings are needed.
Remote command:
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
[SENSe<Ch>:]CORRection:POWer:DATA
[SENSe<Ch>:]CORRection:POWer:DATA:PORT<PhyPt>
Power Cal dialog – Leveling Table
In "Leveling Table" mode, the "Power Cal" dialog guides you through the source power
leveling of a single port.
The dialog shows the selected source port and allows you to define the leveling grid for
the current channel. A circuit diagram visualizes the setup.
●
The leveling function and the "Leveling Table" variant of the "Power Cal" dialog are
only available if option R&S
ZNA-K8 is installed. Their main focus is to enable
source power control for mm-wave converter ports. However, leveling can also be
done for "normal" VNA ports.
●
Leveling requires an existing
●
Although leveling is performed in the standard power cal wizard, the created level-
ing dataset is applied to all channels in the current recall set. A "Preset" does not
affect the leveling dataset and whether or not it is used for power control.
Background information
Refer to
Access:
> "Leveling Table"
Cal softtool