GUI reference
R&S
®
ZNA
478
User Manual 1178.6462.02 ─ 20
Attenuation factors are port- and channel-specific. Possible values are 0
dB to 35
dB in
steps of 5
dB. The analyzer rounds any entered value below the maximum attenuation
to the closest step.
Note:
Receiver step attenuators are optional hardware (see
Remote command:
Preamp. Gain
Sets the gain of the
Internal low power spur reduction amplifier
(b1) and the
(b2)
Remote command:
[SENSe<Ch>:]PAMPlifier2[:STATe]
[SENSe<Ch>:]PAMPlifier2[:VALue
5.8.2.2
Power Config dialog
Except for the ALC settings, which are covered in the
, the "Power
Config" dialog offers the same settings as the
ter 5.12.2.2, "Port Settings dialog"
5.8.2.3
ALC Config dialog
Enables and configures automatic level control (ALC) for the current channel.
Access:
●
Channel – [Pwr
Bw
Avg] > "Power" > "ALC Config..."
●
> "Arbitrary Power" tab > "ALC Params" column
Background information
See
Chapter 4.1.5, "Automatic level control"
Stimulus softtool