GUI reference
R&S
®
ZNA
347
User Manual 1178.6462.02 ─ 20
Compression Point Power In / Compression Point Power Out / Compression
Point S-Param
Selects the input power, output power, or transmission S-parameter value at the
<x
dB> compression point as the measured quantity, where <x
dB> is the specified
Remote command:
'
CALCulate<Ch>:PARameter:MEASure
'<TraceName>',
'CmpPtPin' | 'CmpPtPout' | 'CmpPtS'
CALCulate<Ch>:PARameter:SDEFine
'<TraceName>', 'CmpPtPin' |
'CmpPtPout' | 'CmpPtS'
Compression Value
With a "Compression Value" of
x
, the gain compression measurement searches for the
x
dB compression points of the power sweeps.
The
x
dB compression point is defined as the stimulus level where the response value
has dropped by
x
dB compared to the response value at
.
Remote command:
[SENSe<Ch>:]FREQuency:COMPression:POINt
5.2.7.2
Amplifier Compression setup dialog
The "Amplifier Compression" setup dialog is an instance of the
. It allows you to set up one or more channels for compression point over frequency
measurements on non-frequency converting DUTs.
Access:
> "S-Params" > "Gain Compression" > "Amplifier Compression"
Meas softtool