Concepts and features
R&S
®
ZNA
240
User Manual 1178.6462.02 ─ 20
O U T
O U T
IN
IN
MEAS
SO URCE
REF
O U T
O U T
IN
IN
MEAS
SO URCE
REF
Figure 4-24: Direct access to both ports, coupler as combiner
Intermodulation measurement results
The intermodulation measurement provides two different types of results:
●
In the swept measurement, the analyzer performs a frequency or power sweep of
the two-tone stimulus signal and displays the selected intermodulation quantities
as a function of the lower-tone frequency or power.
●
In the intermodulation spectrum measurement ("CW Mode Spectrum"), the fre-
quency and power of the lower and upper tones is kept constant. The analyzer dis-
plays all intermodulation products near the signals up to a selectable order.