Parameters:
<Offset>
Port-specific power offset
Range:
-300 dB to +300 dB (adjust to the generator's actual
source power range and the test setup)
Increment: 0.01 dB (other values are rounded)
*RST:
0 dB
Default unit: dB
<Mode>
ONLY | CPADd
ONLY - only the port-specific power is used; the channel power
is ignored.
CPADd - the port-specific power is added as an offset to the
channel power.
*RST:
CPADd
Example:
*RST; SOUR:POW -6
Reset the instrument, activating a frequency sweep and set the
internal source power (channel power) for the default channel 1
to -6 dBm.
SYST:COMM:RDEV:GEN1:DEF 'Ext. Gen. 1', 'SME02',
'gpib0', '21';
SYST:COMM:RDEV:GEN2:DEF 'Ext. Gen. 2', 'SME02',
'gpib0', '22'
Configure an R&S SME02 generator as external generator no.
1, assigning the name "Ext. Gen. 1" and a GPIB address 21.
Configure a second R&S SME02 generator, assigning the name
"Ext. Gen. 2" and a GPIB address 22.
SOUR:POW:GEN1:OFFS 6, ONLY; :SOUR:POW:GEN2:OFFS
6, CPAD
Set the source power of generator no. 1 to +6 dBm, the source
power of generator no. 2 to 0 dBm.
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:PERManent[:STATe]
<Boolean>
Defines whether the external generator is available as an external signal source in the
test setup. External generators are always on for all partial measurements.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer. This suffix is ignored; the gen-
erator is selected via
<Gen>
.
<Gen>
Generator number
Parameters:
<Boolean>
ON - generator no.
<PhyPt>
is available.
OFF - generator is available only if a measurement result with a
generator drive port is selected
*RST:
OFF
SCPI command reference