Command reference
R&S
®
ZNA
1389
User Manual 1178.6462.02 ─ 20
Parameters for setting and query:
<ReceiverName>
String parameter defining the wave(s) followed by the drive port;
see examples below and
Example:
*RST; :SENSe:POWer:GAINcontrol:GLOBal MANual
Activate independent GC settings for each drive port.
SENSe:SEGMent1:ADD; :SENSe:SWEep:TYPE SEGMent
Create a new sweep segment no. 1 and activate segmented
sweep type.
SENSe:SEGMent1:POWer:GAINcontrol:CONTrol ON
Enable segmented GC.
SENSe:SEGMent1:POWer:GAINcontrol 'B2D1', LNO
Set the ACG for the received wave b
2
(port 2) in sweep segment
no. 1 to "Low Noise". The setting applies while port 1 is the drive
port.
SENSe:SEGMent1:POWer:GAINcontrol 'A2B2B3D2',
LNO
Set the ACG for the waves a
2
, b
2
, and b
3
in sweep segment no.
1 to "Low Noise". The setting applies while port 2 is the drive
port.
SENSe:SEGMent1:POWer:GAINcontrol? 'B1D2'
Query the ACG setting for the wave b
1
in sweep segment no. 1
while port 2 is the drive port. The response is
1, LDIS
(default
setting).
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol:ALL
<Mode>
For the selected segment, this command applies the same manual gain control (GC)
<Mode> to all a and b wave receivers.
These settings apply if and only if segmented GC is enabled (see
]SEGMent<Seg>:POWer:GAINcontrol:CONTrol
).
Note:
●
This applies to the wideband signal path only.
●
or
[SENSe<Ch>:]POWer:GAINcontrol:ALL
instead, if no distinction between dif-
ferent sweep segments shall be made.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Setting parameters:
<Mode>
LDIStortion | LNOise
LDIStortion
0 dB gain, for high input levels
SCPI command reference