[SENSe<Ch>:]PATH<Pt>:IFRequency
<Numerator>, <Denominator>, <Offset>,
<Sweep type>
Defines the freqency conversion formula for Direct IF Access Input / Output ports IF
Reference <Pt> and IF Meas <Pt> in channel <Ch>.
For <Sweep type>
FB
, the IF frequency f
IF
is calculated according to
f
IF
= <Numerator>/<Denominator> * f
b
+ <Offset>,
where f
b
represents channel base frequency.
For <Sweep Type> CW, f
IF
is calculated according to
f
IF
= <Offset>,
i.e. f
IF
is constant.
Make sure the converted frequency or frequency range is within the Direct IF Access
Input / Output ports' frequency range. If the resulting frequencies are not (completely)
within the allowed range, the analyzer returns an error message. Nevertheless, the fre-
quency formula is applied.
Note:
This setting only takes effect in one of the following cases:
●
[SENSe<Ch>:]PATH<Pt>:REFerence:IFSWitch?
IN
.
●
[SENSe<Ch>:]PATH<Pt>:REFerence:IFSWitch?
OUT
and
[SENSe<Ch>:]PATH<Pt>:IFOutauto
OFF
.
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
Setting parameters:
<Numerator>
The specified numerator is rounded to the next higher integer;
zero is not allowed.
<Denominator>
The specified denominator must be positive. It is rounded to the
next higher integer.
<Offset>
The specified offset value is rounded to (positive or negative)
multiples of 1 Hz.
<Sweep type>
FB | CW