Command reference
R&S
®
ZNA
1268
User Manual 1178.6462.02 ─ 20
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:TERMination
<ReceivingPort>
Acquires receiver noise data in Channel <Ch>, with a matched load connected to
<ReceivingPort>. This is part of a noise figure calibration using a calibration unit..
Suffix:
<Ch>
.
Channel number
Setting parameters:
<ReceivingPort>
Port number of the receiving port
Example:
SENSe1:CORRection:COLLect:AUTO:TERMination 2
Acquire noise data at receive port 2.
Usage:
Setting only
Options:
R&S
ZNA-K30
[SENSe:]CORRection:COLLect:AVERage
<Average>
Activates automatic averaging, which means that the VNA may perform multiple cali-
bration sweeps and apply averaging to reduce trace noise. In contrast to manual aver-
aging (see
) the number of calibration sweeps is
calculated automatically.
Parameters:
<Average>
AUTO | MANual
Manual operation:
See
[SENSe:]CORRection:COLLect:CHANnels:ALL
<Boolean>
Enables calibration of all channels in the active recall set. The command is valid for the
following calibration methods:
●
Manual system error correction
●
Automatic system error correction (cal unit)
●
Manual SMARterCal
●
Automatic SMARTerCal (cal unit)
A scalar power calibration is not affected.
Parameters:
<Boolean>
ON
–
calibrate all channels.
OFF
–
calibrate the active channel only.
Example:
See
[SENSe<Ch>:]CORRection:COLLect:SAVE:
SCPI command reference