Table 3-6: Status bar information - color coding
Color
Type
Description
red
Error
An error occurred at the start or during a measurement, e.g. due to missing
data or wrong settings, so that the measurement cannot be started or com-
pleted correctly.
orange
Warning
An irregular situation occurred during measurement, e.g. the settings no lon-
ger match the displayed results, or the connection to an external device was
interrupted temporarily.
gray
Information
Information on the status of individual processing steps.
no color
No errors
No message displayed - normal operation.
green
Measurement
successful
Some applications visualize that the measurement was successfull by show-
ing a messsage.
If any error information is available for a measurement channel, an exclamation mark is
displayed next to the channel name ( ). This is particularly useful when the MultiView
tab is displayed, as the status bar in the MultiView tab always displays the information
for the currently selected measurement only.
Furthermore, a status bit is set in the
STATus:QUEStionable:EXTended:INFO
reg-
ister for the application concerned (see
"STATus:QUEStionable:EXTended:INFO Reg-
on page 270). Messages of a specific type can be queried using the
SYST:ERR:EXT?
command, see
Table 3-7: Keywords in the status bar
Message
Meaning
IF OVLD
Overload of the IF signal path in the A/D converter or in the digital IF.
Increase attenuation or remove the signal from the RF input.
LOUNL
Error in the instrument's frequency processing hardware was detected.
NO REF
Instrument was set to an external reference but no signal was detected on the refer-
ence input.
OVEN
The OCXO reference frequency has not yet reached its operating temperature. The
message usually disappears a few minutes after power has been switched on.
RF OVLD
Overload of the input mixer or of the analog IF path.
Reduce the input level.
UNCAL
One of the following conditions applies:
●
Correction data has been switched off.
●
No correction values are available, for example after a firmware update.
●
Record the correction data by performing a self alignment
WRONG_FW
The firmware version is out-of-date and does not support the currently installed hard-
ware. Until the firmware version is updated, this error message is displayed and self-
alignment fails.
(For details refer to the R&S
FSPN User Manual).