Remote Control
R&S
®
FSPN
414
User Manual 1179.4363.02 ─ 01
AVERage
The average is formed over several sweeps. The "Sweep/Aver-
age Count" determines the number of averaging procedures.
MAXHold
The maximum value is determined over several sweeps and dis-
played. The R&S
FSPN saves the sweep result in the trace
memory only if the new value is greater than the previous one.
MINHold
The minimum value is determined from several measurements
and displayed. The R&S
FSPN saves the sweep result in the
trace memory only if the new value is lower than the previous
one.
VIEW
The current contents of the trace memory are frozen and dis-
played.
BLANk
Hides the selected trace.
WRHold
The trace is overwritten when new data is available, but only
after all cross-correlation operations defined for a half decade
are done.
*RST:
Depends on the trace.
Example:
INIT:CONT OFF
Switching to single sweep mode.
SWE:COUN 16
Sets the number of measurements to 16.
DISP:TRAC3:MODE WRIT
Selects clear/write mode for trace 3.
INIT;*WAI
Starts the measurement and waits for the end of the measure-
ment.
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:LABel[:STATe]
<State>
This command turns the display of the trace label for a specific trace on and off.
The command can only be applied to active traces.
Suffix:
<n>
.
1..n
<t>
1..n
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF
Common Analysis and Display Functions