Command reference
R&S
®
ZNB/ZNBT
977
User Manual 1173.9163.02 ─ 62
......................................................................................977
CALCulate<Chn>:RIPPle:CONTrol:DOMain
......................................................................977
.......................................................................................978
CALCulate<Chn>:RIPPle:DELete:ALL
..............................................................................979
CALCulate:RIPPle:DISPlay:RESult:ALL[:STATe]
............................................................... 979
CALCulate<Chn>:RIPPle:DISPlay[:STATe]
....................................................................... 979
.......................................................................................980
..........................................................................................980
CALCulate<Chn>:RIPPle:RDOMain:FORMat
....................................................................981
CALCulate<Chn>:RIPPle:SEGMent:COUNt?
....................................................................981
CALCulate<Chn>:RIPPle:SEGMent<Seg>:LIMit
............................................................... 982
CALCulate<Chn>:RIPPle:SEGMent<Seg>:RESult?
...........................................................982
CALCulate<Chn>:RIPPle:SEGMent<Seg>[:STATe]
........................................................... 983
CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:STARt
...............................................983
CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:STOP
............................................... 983
CALCulate<Chn>:RIPPle:SOUNd[:STATe]
........................................................................984
......................................................................................984
CALCulate<Chn>:RIPPle:STATe:AREA
............................................................................ 985
CALCulate<Chn>:RIPPle:CLEar
Resets the limit check results for the ripple test.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Usage:
Event
Manual operation:
See
CALCulate<Chn>:RIPPle:CONTrol:DOMain
<SweepType>
Deletes the existing ripple limit ranges and (re-)defines the physical units of the stimu-
lus values of the ripple limit lines. The unit of the ripple limit is defined via
CALCulate<Chn>:RIPPle:RDOMain:FORMat
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Setting parameters:
<SweepType>
FLIN | FLOG | FSEG | FSINgle | TLIN | TLOG | PLIN | PLOG |
PSINgle
Keywords for the units of the stimulus values; frequency, power,
and time units.
The selected unit must be compatible with the sweep type (see
FLOG, FSEG and FSINgle, s for TLIN and TLOG, dBm for PLIN,
PLOG and PSINgle. Otherwise the ripple limit lines cannot be
displayed and no ripple limit check is possible.
*RST:
FLIN
SCPI command reference