GUI reference
R&S
®
ZNB/ZNBT
555
User Manual 1173.9163.02 ─ 62
●
The measurement parameters are channel-specific. However only one configura-
tion can be measured at a time.
●
Voltage and current measurements on the RFFE and GPIO pins are only possible
with
Var. 03
of the extension board R&S
ZN-B15/Z15 (part number 1323.9355.03
or 1325.5905.03).
Output Voltage
Defines the output voltages for the voltage/current measurements on the RFFE pins.
Note:
The Data and Clock pins always use the same output voltages; their values can-
not be set independently.
The "Output Voltage" and
settings do not take effect until
is
pressed.
Remote command:
CONTrol<Ch>:RFFE<Bus>:TEST:DATA
CONTrol<Ch>:RFFE<Bus>:TEST:CLOCk
CONTrol<Ch>:RFFE<Bus>:TEST:VIO
Range / Shunt
Defines the current range for the voltage and current measurement on the respective
RFFE pin. The analyzer firmware automatically selects a suitable shunt resistance.
and "Range" / "Shunt" settings do not take effect until
is
pressed.
Remote command:
Set
Use the "Set" button to apply all RFFE configuration changes and to activate the con-
figured voltage and
Remote command: