GUI reference
R&S
®
ZNB/ZNBT
476
User Manual 1173.9163.02 ─ 62
functionality is the same as described for the "SingleCal" version of the wizard (see
"Step 2: Connectors and Cal Kits"
Back
Go back to
.
Start
Start
.
Step 3: Calibration
Allows you to acquire error correction data for every required port (pair) and calibration
standard, where "required" depends on the selected ports and calibration types.
Same logic as in the single calibration version of the wizard (see
The possibility to "share" measurements between the calibrations and the "Reduced
Through" logic help to keep the number of measurements as small as possible (see
Chapter 4.5.1.11, "Full n-Port calibration with reduced number of Through connec-
tions"
5.11.1.7
Power Cal wizard
The "Power Cal" wizard displays a single screen, showing the current source and
receive ports. It allows you to perform "Ref. Receiver", "Meas. Receiver" and scalar
source power ("Source Flatness") calibrations, based on the current power calibration
settings (see
Chapter 5.11.3, "Pwr Cal Settings tab"
Access:
Channel – [Cal] > "Start Cal" > "Power Cal..."
Cal softtool