Command reference
R&S
®
ZNB/ZNBT
1424
User Manual 1173.9163.02 ─ 62
The corresponding cal group file (
<PresetUserCal>.cal
) must be available in the
cal pool directory
C:\Users\Public\Documents\Rohde-Schwarz\Vna\Calibration\Data\
.
Parameters:
<PresetUserCal>
Name of the cal group (i.e. the name of the cal group file without
path and extension).
Use the empty string to restore the default behavior (no preset
user cal).
Example:
MMEMORY:STORE:CORRection 1, 'Refl1
Trans123.cal'
Adds the active user calibration to the memory pool
SYSTem:PRESet:USER:CAL 'Refl1 Trans123'
Sets the newly created cal pool member as the preset user cal.
SYSTem:PRESet:USER:CAL?
Queries for the preset user cal. Returns
'Refl1 Trans123'
SYSTem:PRESet:USER:CAL ''
Restores the default behaviour of the instrument: no preset user
cal.
Manual operation:
See
SYSTem:PRESet:USER:NAME
<RecallSetFile>
Specifies the name of a recall set file (
*.znx|*.znxml
) to be used for a user-defined
preset.
Parameters:
<RecallSetFile>
String parameter to specify the absolute or relative path of the
recall set file to be loaded. Relative paths are evaluated relative
to the current directory (see
Example:
See
Manual operation:
See
SYSTem:PRESet:USER[:STATe]
<Boolean>
Determines whether a factory or a user-defined preset is performed.
This setting applies to the analyzer GUI and to the remote commands
*RST
and
is set to
ON
.
Parameters:
<Boolean>
OFF (0)
A factory preset is performed.
ON (1)
A user-defined preset is performed, if the setup file selected
using
erwise, a factory preset is performed.