Command reference
R&S
®
ZNB/ZNBT
1400
User Manual 1173.9163.02 ─ 62
Example:
See
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:
Usage:
Event
Manual operation:
See
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:SPCorrection[:STATe]
<Boolean>
Gets/sets the state of the built-in S-parameter correction that is available on certain
R&S®NRP-Z power sensors.
Note that this state is persistently stored on the power sensor (and NOT on the
R&S
ZNB/ZNBT).
See Application Note 1GP70 "Using S-Parameters with R&S®NRP-Z Power Sensors"
for background information. This Application Note is available on the Rohde & Schwarz
internet at
http://www.rohde-schwarz.com/appnotes/1GP70
Suffix:
<Pmtr>
.
Number of the configured power meter.
Parameters:
<Boolean>
ON | OFF
Manual operation:
See
"Deembed Two-Port (All Channels)"
SYSTem:COMMunicate:RDEVice:SMATrix<Matr>:CATalog?
Queries the numbers of the switch matrices currently part of the RF configuration (see
on page 1135). The response is a string containing a comma-
separated list of switch matrix numbers.
Suffix:
<Matr>
.
This suffix is ignored and can be omitted.
Example:
Suppose switch matrices 1 and 2 are currently connected to the
R&S
ZNB/ZNBT.
SYST:COMM:RDEV:SMAT1:CAT?
Query the numbers of the connected switch matrices. The
response is '1, 2'.
SYST:COMM:RDEV:SMAT1:COUN?
Query the number of connected switch matrices. The response
is 2.
Usage:
Query only
SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:ABORt
Aborts a manual RF connection configuration.
SCPI command reference