Command reference
R&S
®
ZNB/ZNBT
1312
User Manual 1173.9163.02 ─ 62
●
or
[SENSe<Ch>:]POWer:GAINcontrol:ALL
instead, if no distinction between dif-
ferent sweep segments shall be made.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Setting parameters:
<Mode>
AUTO | LDIStortion | LNOise
AUTO
automatic/adaptive gain control (AGC) according to the RF input
level
LDIStortion
fixed, small IF gain, for high input levels
LNOise
fixed, large IF gain, for low input levels
*RST:
LDIStortion
Example:
*RST; :SENSe:POWer:GAINcontrol:GLOBal MANual
Enable the manual configuration of independent GC settings for
each drive port.
:SENSe:SEGMent1:ADD; :SENSe:SWEep:TYPE SEGMent
Create a new sweep segment no. 1 and activate segmented
sweep type.
SENSe:SEGMent1:POWer:GAINcontrol:CONTrol ON
Enable segmented GC.
SENSe:SEGMent1:POWer:GAINcontrol:ALL LNO
For sweep segment no. 1, set the GC mode of all a and b wave
receivers to "Low Noise". Segmented GC is enabled automati-
cally.
Usage:
Setting only
Firmware/software:
V2.75 or higher
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol:CONTrol
<Boolean>
Defines whether common or independent gain control (GC) settings shall be used for
the individual sweep segments.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. This suffix is ignored.
SCPI command reference