Operating Manual 1302.6163.12 - 03
4.37
R&S ESU
Receiver Mode
MARGIN
The
MARGIN
softkey activates the entry field of the margin, i.e. of an additional
acceptance threshold for the determination of the peak list. The limit line currently
used is shifted by this amount for defining the maxima. The range of values is -200
dB to 200 dB.
Remote command:
:CALC:PEAK:MARG –200dB...200dB
FINAL
MEAS TIME
The
FINAL MEAS TIME
softkey activates the entry field of the time of final measure-
ment.
Remote command:
FME:TIME <num_value>
AUTOMATIC
FINAL
The
AUTOMATIC FINAL
softkey activates the automatic mode for the final mea-
surement, i.e. a final measurement will be performed automatically and immediately
each time a value out of limit is detected.
Remote command:
:FME:AUTO ON
INTER
ACTIVE
The
INTERACTIVE
softkey selects the following sequence for the final measure-
ment:
• The prescan is interrupted –
HOLD SCAN
state.
• The bar graph measurement is started in the free running mode.
• The signal can be exactly analyzed by modifying the receiver settings.
• The actual final measurement is started, the receiver settings being restored
except the frequency.
• The current frequency replaces the original one in the frequency list (drifting
interference sources).
• The prescan is continued at the frequency at which it was interrupted.
Remote command:
:FME:AUTO OFF
RUN
SCAN
The
RUN SCAN
softkey starts the prescan. The
HOLD SCAN
submenu is called.
For information about
RUN SCAN
Out-of-limit value for
AUTOMATIC FINAL
If an out-of-limit value is detected, the R&S ESU automatically goes to the HOLD
SCAN state and starts the final measurement. The
submenu comes up
.
On completion of the final measure-
ment, the R&S ESU continues the prescan, and the
HOLD SCAN
submenu is dis-
played again.
With the
AUTOMATIC FINAL
softkey in the CONTINUE FINAL MEAS submenu a
switchover can be made to the automatic mode before the measurement is
started anew.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...