R&S ESU
Remote Control – Description of Commands
DIAGnostic Subsystem
6.100
Operating Manual 1302.6163.12 - 03
6.7 DIAGnostic Subsystem
The DIAGnostic subsystem contains the commands which support instrument diagnostics for mainte-
nance, service and repair. In accordance with the SCPI standard, all of these commands are device-spe-
cific.
The measurement windows are selected by DIAGnostic1 (screen A) or DIAGnostic2 (screen B).
DIAGnostic<1|2>:SERVice:INPut[:SELect]
CALibration | RF
This command toggles between the RF input on the front panel and the internal 128 MHz reference
signal in the selected measurement window. The level of the 128-MHz signals can be selected by
DIAGnostic<1|2>:SERVice:CSOource[:POWer]
command.
Example:
"DIAG:SERV:INP CAL"
’ Selects the internal reference signal.
Characteristics:
*RST value: RF
SCPI: device-specific
Mode:
R, A
DIAGnostic<1|2>:SERVice:INPut:PULSed[:STATe]
ON | OFF
This command toggles the calibration signal in the selected measurement window between pulsed and
non-pulsed.
The selection takes effect only if the RF input has been set to the internal reference signal using the
DIAGnostic<1|2>:SERVice:INPut[:SELect]
command.
Example:
"DIAG:SERV:INP CAL"
"DIAG:SERV:INP:PULS ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
DIAGnostic<1|2>:SERVice:INPut:PULSed:PRATe
10 kHz | 62.5 kHz | 1 MHz | 128 MHz | 640 MHz
This command selects the pulse rate for the pulsed calibration signal in the selected measurement
window.
Example:
"DIAG:SERV:INP:PULS:PRAT 128 MHz"
Characteristics:
*RST value: 128 MHz
SCPI: device-specific
Mode:
R, A
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...