R&S ESU
Receiver Mode
4.16
Operating Manual 1302.6163.12 - 03
4.5.2
Level Display and RF Input Configuration – AMPT
Key
The
AMPT
key is used to set the input attenuation, the
, the auto range
function and the
.
In addition, the level display range for the scan can be set.
AMPT
*) LN PREAMP ON OFF is only available with option R&S ESU-B24
RF ATTEN
MANUAL
The
RF ATTEN MANUAL
softkey activates the attenuation entry field.
The attenuation can be set between 0 and 75 dB in 5 dB steps. Other entries are
rounded up to the next higher integer.
Remote command:
INP:ATT 40 DB
Preamplifier
R&S ESU is provided with a switchable preamplifier of 20 dB gain in the frequency
range up to 3.6 GHz. An additional low noise preamplifier which covers the whole
frequency range is available as option R&S ESU-B24.
Switching on the preamplifier reduces the total mark figure of the R&S ESU and thus
improves the sensitivity.
The disadvantage of a poorer large-signal immunity (intermodulation) is reduced by
the connected preselector. The signal level of the subsequent mixer is 20 dB higher
so that the maximum input level is reduced by the gain of the preamplifier. The use
of the preamplifier is recommended when measurements with a maximum sensitiv-
ity are to be performed. On the other hand, if the measurement should be performed
at maximum dynamic range, the preamplifier should be switched off.
GRID LEVEL
Side menu
dBmV | dBm | dBmA | dBpW |
dBpT | dBmV | dB*/MHz
To protect the input mixer against inadvertent overload, 0 dB can only be
switched on when the
10 dB MIN
softkey is switched off.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...