Operating Manual 1302.6163.12 - 03
5.25
R&S ESU
Remote Control – Basics
Status Reporting System
5.7.3.6
STATus:QUEStionable:ACPLimit Register
This register comprises information about the observance of limits during adjacent
power measurements. It can be queried with commands
STATus:QUEStionable:
ACPLimit:CONDition?
and
STATus:QUEStionable:ACPLimit[:EVENt]?
.
9
LIMit
(device-specific)
This bit is set if a limit value is violated (see also section
“STATus:QUEStionable:LIMit<1|2>
10
LMARgin
This bit is set if a margin is violated (see also section
“STATus:QUEStionable:LMARgin<1|2>
11
This bit is not used
12
ACPLimit
This bit is set if a limit for the adjacent channel power measurement is violated (see also
section
“STATus:QUEStionable:ACPLimit Register” on page 5.25
)
13
TRANSducer break
14
This bit is not used
15
This bit is always 0.
Bit
No.
Meaning
Bit
No.
Meaning
0
ADJ UPPer FAIL(Screen A)
This bit is set if in screen A. the limit is exceeded in the upper adjacent channel
1
ADJ LOWer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the lower adjacent channel.
2
ALT1 UPPer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the upper 1st alternate channel.
3
ALT1 LOWer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the lower 1st alternate channel.
4
ALT2 UPPer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the upper 2nd alternate channel.
5
ALT2 LOWer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the lower 2nd alternate channel.
6
ALT3 to 11 LOWer/UPPer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in one of the lower or upper alternate channels
3 to 11.
7
not used
8
ADJ UPPer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the upper adjacent channel.
9
ADJ LOWer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the lower adjacent channel.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...