R&S ESU
Instrument Functions
Tracking Generator – Option R&S FSU-B9
4.332
Operating Manual 1302.6163.12 - 03
4.10.7
Power Offset of the Tracking Generator
POWER
SWEEP
The
POWER SWEEP
softkey opens a submenu for activates or deactivates the
power sweep.
POWER SWP
ON/OFF
The
POWER SWP ON/OFF
softkey activates or deactivates the power sweep. If the
power sweep is ON the enhancement label TGPWR is shown and the EMI Test
Receiver is set in zero span mode (span = 0Hz). During the sweep time of the zero
span the power at the internal tracking generator is changed linear from start power
to stop power. The start and stop power values are shown on the right side below
the diagram.
Remote command:
:SOUR:POW:MODE SWE
:SOUR:POW:MODE FIX
START POWER
The
START POWER
softkey defines the start power of the power sweep.
The start power can be set between -30 dBm and +5 dBm.
With the option R&S FSU-B12 the power can be set between -100 and + 5 dBm.
Remote command:
:SOUR:POW:STAR –20dBm
STOP POWER
The
STOP POWER
softkey defines the stop power of the power sweep.
The start power can be set between -30 dBm and +5 dBm.
With the option R&S FSU-B12 the power can be set between -100 and + 5 dBm.
The stop value can be smaller than the start value.
Remote command:
:SOUR:POW:STOP –10dBm
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...