Operating Manual 1302.6163.12 - 03
4.265
R&S ESU
Instrument Functions
Basic Settings
4.9.3.2
Control of V-Networks (LISNs)
LISN
The
LISN
softkey opens the submenu including the settings required to control V-
networks (LISNs).
ESH2-Z5 /
ESH3-Z5 /
ENV 4200 /
ENV 216 / OFF
The
ESH2-Z5, ESH3-Z5
,
ENV 4200
,
ENV 216
and
OFF
softkeys select the V-net-
work to be controlled via the user port. They are toggle softkeys, and only one of
them can be activated at a time.
Remote command:
INP:LISN TWOP | FOUR | ENV4200 | ENV216 |
OFF
PHASE N /
PHASE L1 /
PHASE L2 /
PHASE L3
The
PHASE N
,
PHASE L1
,
PHASE L2
, and
PHASE L3
softkeys select the phase of
the V-network on which the RFI voltage is to be measured.
Remote command:
INP:LISN:PHAS L1|L2|L3|N
150 KHZ
HIGHPASS
The
150 KHZ HIGHPASS
softkey switches an 150 kHz highpass on to protect the
receiver input from high signal levels below 150 kHz (only for ENV 216).
Remote command:
INP:LISN:FILT:HPAS ON | OFF
PE GROUNDED
/
PE FLOATING
The
PE GROUNDED
and
PE FLOATING
softkeys switch the protective earth con-
ductor chokes on or off.
Remote command:
INP:LISN:PEAR GRO|FLO
ESH2-Z5 / ESH3-Z5 / ENV 4200 / ENV 216 / OFF
PHASE N / PHASE L1 / PHASE L2 / PHASE L3
ESH2-Z5
four-line V-network is controlled.
ESH3-Z5
two-line V-network is controlled.
ENV 4200
four-line V-network is controlled.
ENV 216
two-line V-network is controlled.
OFF
remote control is deactivated.
PHASE N
RFI on phase N is measured.
PHASE L1
RFI on phase L1 is measured.
PHASE L2
RFI on phase L2 is measured
(only for ESH2-Z5/ENV 4200).
PHASE L3
RFI on phase L3 is measured
(only for ESH2-Z5/ENV 4200).
PE GROUNDED
Protective
earth
conductor choke is switched off.
PE FLOATING
Protective
earth
conductor choke is switched on.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...