All real amplifiers also generate unwanted RF products outside the assigned fre-
quency band. These spurious emissions are usually measured across a wide fre-
quency range, for example from 9 kHz to 7 GHz. The analyzer settings are specified
for each frequency range.
SPURIOUS
EMISSIONS
In the Spurious Emissions measurement mode, the R&S ESU performs measure-
ments in predefined frequency ranges with settings that can be specified individually
for each of these ranges.
For this purpose, the SWEEP TABLE settings or the current device settings are
used. Up to 20 subranges can be defined (they need not directly follow one another)
across which the R&S ESU sweeps in subsequent order. However, the measure-
ment ranges must not overlap. The measurement parameters can be selected inde-
pendently from each other in every subrange (
SWEEP LIST
menu, EDIT SWEEP
LIST).
!
!
Side menu
!
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...