Operating Manual 1302.6163.12 - 03
4.223
R&S ESU
Instrument Functions
Analyzer Mode
DECIM SEP
In the case of floating-point numbers, use the
DECIM SEP
softkey to select between
'.' (decimal point) and ',' (comma) as the decimal delimiter for the ASCII FILE
EXPORT function.
By selecting the decimal delimiter, various language versions of evaluation pro-
grams (e.g. MS Excel) are supported.
Remote command:
FORM:DEXP:DSEP POIN
PAGE UP /
PAGE DOWN
Using
PAGE UP
and
PAGE DOWN
, you can scroll forward and backward through
the peak list pages.
They are active only as long as a peak list is displayed.
4.8.13.10
Spectrum Emission Mask Measurement
The Spectrum Emission Mask (SEM) measurement defines a measurement that
monitors compliance with a spectral mask. The SEM measurement of the base unit
allows a flexible definition of all parameters in the SEM measurement. The analyzer
performs measurements in predefined frequency ranges with settings that can be
specified individually for each of these ranges.
For this purpose, the SWEEP LIST settings or the current instrument settings are
used. Up to 20 subranges can be defined (they need not directly follow one another)
across which the analyzer sweeps in subsequent order. However, the measurement
ranges must not overlap. The measurement parameters can be selected indepen-
dently from each other in every subrange (SWEEP LIST menu, EDIT SWEEP LIST).
In the Spectrum Emission Mask, as in the Spurious Emissions measurement, a
range defines a segment, for which you can define the following parameters sepa-
rately: start and stop frequency, RBW, VBW, sweep time, reference level, attenuator
settings, and limit values. Via the sweep list, you define the ranges and their settings
(for details on settings refer to the SWEEP LIST softkey).
The following rules apply to ranges:
• The minimum span of a range is 10 Hz.
• The individual ranges must not overlap (but need not directly follow one another).
• The maximum number of ranges is 20.
• A minimum of three ranges is mandatory.
• The reference range cannot be deleted (it is marked in with "Ref").
The frequency range where measurements are actually performed is set by the start
and stop frequency parameters of the analyzer; these parameters are independent
of the sweep ranges. It is thus possible to define sweep ranges for a measurement
task that can be stored and reloaded and to quickly and easily set the frequency
range to be actually measured by means of two parameters; complex editing in the
sweep table is not necessary.
Provided XML Files for the Spectrum Emission Mask Measurement
You can change the settings manually or via XML files. The XML files offer a quick
way to change the configuration. A set of ready-made XML files for different stan-
dards is already provided. For details see Table1: Provided XML files. You can also
create and use your own XML files (for details see
“Format Description of Spectrum
Emission Mask XML files” on page 4.228
). All XML files are stored under D:
\r_s\instr\sem_std. Use the LOAD STANDARD softkey for quick access to the avail-
able XML files.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...