Instrument Functions
Analyzer Mode
4.214
Operating Manual 1302.6163.12 - 03
TOI MKR CALC
SRCH
The TOI measurement is based on positioning the markers 3 and 4 onto the position
of the calculated intermodulation products. The accuracy of these frequency calcula-
tions depend on the precision of the two carrier frequencies found with peak search.
The softkey
TOI MKR CALC SRCH
allows to switch between the calculated marker
position (CALC) and perfoming a local peak search near the expected frequencies
(SRCH) instead of the faster position calculation.
By default, the faster CALC method is used.
Remote command:
CALC:MARK:FUNC:TOI:MARK CALC
4.8.13.8
Harmonic Measurement
HARMONIC
DISTOR
The
HARMONIC DISTOR
softkey opens this submenu and activates the harmonic
distortion measurement.
In the upper screen the zero span sweeps on all harmonics are shown, separated by
a grid line. This allows a very good overview about the measurement. In the lower
screen the mean RMS results are displayed in numerical values and the THD values
can be seen in the marker info field.
The resolution bandwidth will be automatically adjusted: RBW
n
= RBW
1
* n, if that
RBW is not available the next higher value is used.
The results can be obtained via the following remote commands:
Trace read out via the normal trace subsystem. The first harmonic frequency can be
read out via the center frequency command.
THD value comma separated in % and dB:
Remote command:
CALC:MARK:FUNC:HARM:DIST? TOT
Comma separated list of harmonic levels, for each harmonic one value:
Remote command:
CALC:MARK:FUNC:HARM:LIST?
HARMONIC
ON | OFF
The
HARMONIC ON/OFF
softkey activates the harmonic distortion measurement.
With this measurement it is possible to measure easily the harmonics e.g. from a
VCO. In addition the THD (total harmonic distortion) is calculated in % and dB.
There are 2 possible modes within the harmonic measurement. When entering the
harmonic measurement from a frequency sweep (span > 0 Hz) an automatic search
for the first harmonic is carried out within that given frequency range. Also a level
adjust will be carried out. Is the zero span mode active before entering the harmonic
measurement, the center frequency is unchanged.
Remote command:
CALC:MARK:FUNC:HARM:STAT ON | OFF
The results can be obtained via the following remote commands:
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...