Operating Manual 1302.6163.12 - 03
4.143
R&S ESU
Instrument Functions
Analyzer Mode
4.8.9
Recording the Correction Data – CAL Key
The R&S ESU obtains its high measurement accuracy through its inbuilt self-align-
ment method.
The correction data and characteristics required for the alignment are determined by
comparison of the results at different settings with the known characteristics of the
high-precision calibration signal source of R&S ESU at 128 MHz. The correction
data are then available in the instrument as a file and can be displayed by means of
the
softkey.
For service purposes the use of correction data can be deactivated by means of the
CAL CORR ON/OFF
softkey. If the correction data recording is aborted, the last
complete correction data set is restored.
The
CAL
key opens a menu with the available functions for recording, displaying
and activating the data for self alignment.
CAL
CAL TOTAL
The
CAL TOTAL
softkey starts the recording of correction data of the instrument.
If the correction data recording has failed or if the correction values are deactivated
(
CAL CORR = OFF
softkey),
UNCAL
is displayed in the status field.
Remote command:
*CAL?
CAL ABORT
The
CAL ABORT
softkey interrupts the recording of correction data and restores the
last complete correction data set.
Remote command:
CAL:ABOR
CAL CORR
ON | OFF
The
CAL CORR ON/OFF
softkey switches the calibration data on/off.
• ON: The status message depends upon the results of the total calibration.
• OFF: The message
UNCAL
appears in the status line.
Remote command:
CAL:STAT ON
YIG CORR
ON | OFF
The
YIG CORR ON/OFF
softkey switches on or off the automatic, cyclic correction
of the temperature-dependent frequency drift of the YIG filter.
The term "Calibration" formerly used for the integrated self alignment was often
mistaken for the "true" calibration of the instrument at the test set in production and
in service. It is therefore no longer used although it appears in the abbreviated
form in the name of keys ("CAL...").
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...