4.109
R&S ESU
Instrument Functions
Analyzer Mode
The
UNIT
softkey opens a submenu to select the unit for the level axis.
The default setting is dBm.
In general, the R&S ESU measures the signal voltage at the RF input. The level dis-
play is calibrated in rms values of an unmodulated sinewave signal. In the default
state, the level is displayed at a power of 1 mW (= dBm). Via the known input resis-
tance of 50
Ω
or 75W, conversion to other units is possible. The units dBm, dBmV,
dBµV, dBµA, dBpW, V, A and W are directly convertible.
Remote command:
CALC:UNIT:POW DBM
RF INPUT
AC | DC
The
RF INPUT AC/DC
softkey toggles the RF input of the R&S ESU between AC
and DC coupling.
Remote command:
INP:COUP AC
RF ATTEN
MANUAL
The
RF ATTEN MANUAL
softkey allows the attenuation to be entered irrespective of
the reference level.
The attenuation can be set in 5 dB steps between 0 and 75 dB.
Other entries will be rounded to the next higher integer value.
If the defined reference level cannot be set for the given RF attenuation, the refer-
ence level will be adjusted accordingly and the warning "Limit reached" will be out-
put.
RF ATTEN MANUAL
is the default setting.
Remote command:
INP:ATT 40 DB
RF ATTEN
AUTO
The
RF ATTEN AUTO
softkey sets the RF attenuation automatically as a function of
the selected reference level.
This ensures that the optimum RF attenuation desired by the user is always used.
Remote command:
INP:ATT:AUTO ON
10 dB MIN
ON | OFF
The
10 dB MIN ON/OFF
softkey determines whether or not the 0 dB position of the
attenuator is used when the attenuation is set manually or automatically.
The default setting is
ON
. This means the R&S ESU always leaves at least 10 dB
RF attenuation on in order to protect the input mixer.
The 0 dB position cannot be switched on manually. This prevents that 0 dB are set
by mistake especially for measurements on objects with high signal voltage.
Remote command:
INP:ATT:PROT ON
Side menu
dBpT
The value 0 dB cannot be switched on unless the
softkey is
set to
OFF
in order to protect the input mixer against accidental overload.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...