R&S ESU
Instrument Functions
Analyzer Mode
4.104
Operating Manual 1302.6163.12 - 03
MANUAL
The
MANUAL
softkey activates the window for entering a fixed step size.
Remote command:
FREQ:CENT:STEP 120MHz
Softkeys in time domain:
0.1 * RBW
The
0.1 * RBW
softkey sets the step size for the center frequency entry to 10% of
the resolution bandwidth.
AUTO 0.1 * RBW
corresponds to the default setting.
Remote command:
FREQ:CENT:STEP:LINK RBW
FREQ:CENT:STEP:LINK:FACT 10PCT
0.5 * RBW
The
0.5 * RBW
softkey sets the step size for the center frequency entry to 50% of
the resolution bandwidth.
Remote command:
FREQ:CENT:STEP:LINK RBW
FREQ:CENT:STEP:LINK:FACT 50PCT
X * RBW
The
X * RBW
softkey allows the factor defining the center frequency step size to be
entered as % of the resolution bandwidth.
Values between 1 and 100% in steps of 1% are allowed. The default setting is 10%.
Remote command:
FREQ:CENT:STEP:LINK RBW
FREQ:CENT:STEP:LINK:FACT 20PCT
= CENTER
The
= CENTER
softkey sets the step size coupling to MANUAL and the step size to
the value of the center frequency. This function is especially useful during measure-
ments of the signal harmonic content because by entering the center frequency
each stroke of the
STEP
key selects the center frequency of another harmonic.
Remote command:
--
= MARKER
The
= MARKER
softkey sets the step size coupling to MANUAL and the step size to
the value of the marker. This function is especially useful during measurements of
the signal harmonic content at the marker position because by entering the center
frequency each stroke of the
STEP
key selects the center frequency of another har-
monic.
Remote command:
--
MANUAL
The
MANUAL
softkey activates the window for entering a fixed step size.
Remote command:
FREQ:CENT:STEP 120MHz
START
The
START
softkey activates the window for manually entering the start frequency.
The allowed range of values for the start frequency is:
0 Hz
≤
f
start
≤
f
max
- minspan
Remote command:
FREQ:STAR 20MHz
f
start
start
frequency
minspan smallest selectable span (10 Hz)
f
max
max.
frequency
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...