The delta marker1 x-value can not be changed away from 0 as long as the link func-
tionality is active.
Remote command:
CALCulate<1|2>:DELTamarker<1...4>:LINK
ON | OFF
The suffix at
DELTamarker
can only be 1 or not present, because the
functionality is only available for marker1.
4.6.3
Marker Functions – MKR FCTN Key
In spectrum analysis mode the
MKR FCTN
menu offers further measurements with
the markers. In IF analysis the functionality is limited.
Select the spectrum analysis mode for the following functions:
• Measurement of noise density (
NOISE MEAS
softkey)
• Measurement of phase noise (
PHASE NOISE
softkey)
• Measurement of filter or signal bandwidth (
N DB DOWN
softkey)
• Activating of AF demodulation (
MARKER DEMOD
softkey)
On calling the menu, the entry for the last active marker is activated (
SELECT
MARKER
softkey); if no marker is activated, marker 1 is activated and a maximum
search (
PEAK
softkey) is performed. The marker can be set to the desired trace by
means of the
MKR->TRACE
softkey.
MKR FCTN
4.6.3.1
Activating the Markers
SELECT
MARKER
The
SELECT MARKER
softkey activates the numerical selection of the marker in
the data entry field.
For further information refer to the descriptions in section
PEAK
The
PEAK
softkey sets the active marker/delta marker to the peak of the trace.
For further information refer to the descriptions in section
.
4.6.3.2
Selecting the Trace
MKR->TRACE
The
MKR
−>
TRACE
softkey places the marker on a new trace.
For further information refer to the descriptions in section
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...