Rainin Pipet Lite Series User Manual Download Page 8

6

4. Wipe any excess liquid from the outside of the tip

without touching the orifice.

5. To dispense, rest the end of the tip against the

vessel wall and press the plunger to the first stop.
Hold this position a few seconds, or long enough
for the liquid column to reach equilibrium again.

6. Remove the tip from the receiving vessel without

blowing out the remaining liquid.

7. Return excess sample in the tip to the original

sample container, if desired. Discard the used tip.

Pipetting Liquids of Varying Density

Pipet-Lite lets you compensate for solutions of densi-
ty much different from water, by setting the volume
slightly higher or lower than that required. The com-
pensation amount must be determined empirically.

For example, if pipetting 10 µL of CsCl solution, you
determine that the volume delivered is actually 8.5 µL
(

5 samples). Change the volume setting to 11.8 µL

and repeat the measurements. If the volumes deliv-
ered are still not close enough to 10 µL, make anoth-
er slight volume adjustment until the measurements
are as desired.

Temperature Considerations

Warm or cold liquids can be measured with good pre-
cision by using a consistent pipetting rhythm. This will
help minimize any differences in heating or cooling
effects within the pipette. 

Use a new tip each time for best accuracy and preci-
sion when measuring samples with temperatures
greatly different from ambient, and do not pre-rinse.
As with any air-displacement pipette, you will get best
results if there is no delay between picking up the
sample and dispensing it.

Autoclaving

Autoclavable parts of Pipet-Lite are the shaft and the tip
ejector: 121°C, 1 bar, 15-20 minutes.

Do not autoclave the complete pipette or any
parts other than the shaft and the tip ejector.

9920-237 Rev C GENERIC  9/13/00  3:16 PM  Page 8

Summary of Contents for Pipet Lite Series

Page 1: ...ntinuously adjustable digital microliter pipette 9 pipettes for volume ranges from 0 1 L to 10 mL Pipet Lite 200 L shown pipet lite Magnetic Assist Pipette 9920 237 Rev C GENERIC 9 13 00 3 16 PM Page...

Page 2: ...everse Mode Pipetting 5 Pipetting Liquids of Varying Density 6 Temperature Considerations 6 Autoclaving 6 Tip Ejector Arm Removal 7 Pipet Lite Storage 7 Troubleshooting and Repairs 8 Service Calibrati...

Page 3: ...uctivity polymer Shaft Disposable tip Introduction Pipet Lite is an air displacement pipette which incor porates a major ergonomic improvement to reduce the risk of repetitive strain injury and pain A...

Page 4: ...nths hundredths of mL 10 mL Red mL Black tenths of mL 3 To eliminate errors due to mechanical backlash when setting the desired volume first turn the knob 1 3 turn above the desired volume Then turn t...

Page 5: ...Using such a filter is particularly important when pipetting large vol umes Replace the filter if it gets wet Figure 2 Filter Orientation For 5000 L pipettes insert the small diameter into the shaft...

Page 6: ...actual samples 1 Set the desired volume as described on page 2 2 Attach a new Rainin tip Press the shaft into the tip with only sufficient force to make a good seal 3 Press the plunger button to the...

Page 7: ...ST STOP 4 Consistent immersion depth See table on page 4 5 Pipette vertically or within 20 of vertical 6 Don t invert or lay the pipette flat with liquid in the tip Pre Rinsing Recommended Some soluti...

Page 8: ...ume delivered is actually 8 5 L 5 samples Change the volume setting to 11 8 L and repeat the measurements If the volumes deliv ered are still not close enough to 10 L make anoth er slight volume adjus...

Page 9: ...r arm snaps in place Pipet Lite Storage After use store the pipette in a clean safe place Pipet Lite is a precision instrument and should be treated with the level of care appropriate for laborato ry...

Page 10: ...a small amount of grease to the seal 4 If piston corrosion or staining is evident do not use the instrument Call 800 662 7027 for Service Leaks Inaccuracy Abnormal Stroke 1 Loose shaft Tighten couplin...

Page 11: ...is St Emeryville CA 94608 USA Tel 800 662 7027 Fax 510 652 4403 Rainin Service Center Rainin Road Woburn MA 01801 USA Tel 800 662 7027 Fax 781 935 7631 Rainin Europe Nieuwenhuizenweg 9 2314 XP Leiden...

Page 12: ...2 0 0 024 6 0 0 012 1 0 2 7 0 027 1 3 0 013 2 0 1 5 0 030 0 7 0 014 10 L 1 0 0 02 2 5 0 025 1 2 0 012 5 0 1 5 0 075 0 6 0 03 10 0 1 0 0 1 0 4 0 04 20 L 2 0 02 7 5 0 15 2 0 0 04 5 3 0 0 15 0 9 0 045 10...

Page 13: ...140 6200 145 6200 147 6200 157 6200 160 6200 169 C Short Shaft n a n a 6200 382 6200 383 n a D 6200 133 6200 133 6200 144 6200 148 6200 156 6200 163 6200 168 D Short Tip Ejector n a n a 6200 384 6200...

Page 14: ...B C D E F G H J A B C D E F G H J A B C D E F G H J Legend for Pipet Lite 2 L to 2000 L Models this page and top of next page A Plunger Button B Shaft Coupling C Shaft D Tip Ejector E Piston Assembly...

Page 15: ...G Pipet Lite 5000 L Pipet Lite 10 mL A C H H D E F G A C D E F G Legend for Pipet Lite 5000 L and 10 mL below A Plunger Button C Tip Ejector Arm D Piston Assembly E Piston O ring F Cylinder G Cylinder...

Page 16: ...nin com Pipette Service Phone 800 662 7027 Fax 781 935 7631 E mail service rainin com Direct Order Line Phone 800 472 4646 Fax 781 938 1152 E mail pipets rainin com Web www rainin com From outside the...

Reviews: