continuously-adjustable digitalmicroliterpipette
9 pipettes for volume rangesfrom0.1 µL to 10 mL
Pipet-Lite 200 µL shown
pipet
•
lite
™
Magnetic Assist Pipette
9920-237 Rev C GENERIC 9/13/00 3:16 PM Page 1
Page 1: ...ntinuously adjustable digital microliter pipette 9 pipettes for volume ranges from 0 1 L to 10 mL Pipet Lite 200 L shown pipet lite Magnetic Assist Pipette 9920 237 Rev C GENERIC 9 13 00 3 16 PM Page...
Page 2: ...everse Mode Pipetting 5 Pipetting Liquids of Varying Density 6 Temperature Considerations 6 Autoclaving 6 Tip Ejector Arm Removal 7 Pipet Lite Storage 7 Troubleshooting and Repairs 8 Service Calibrati...
Page 3: ...uctivity polymer Shaft Disposable tip Introduction Pipet Lite is an air displacement pipette which incor porates a major ergonomic improvement to reduce the risk of repetitive strain injury and pain A...
Page 4: ...nths hundredths of mL 10 mL Red mL Black tenths of mL 3 To eliminate errors due to mechanical backlash when setting the desired volume first turn the knob 1 3 turn above the desired volume Then turn t...
Page 5: ...Using such a filter is particularly important when pipetting large vol umes Replace the filter if it gets wet Figure 2 Filter Orientation For 5000 L pipettes insert the small diameter into the shaft...
Page 6: ...actual samples 1 Set the desired volume as described on page 2 2 Attach a new Rainin tip Press the shaft into the tip with only sufficient force to make a good seal 3 Press the plunger button to the...
Page 7: ...ST STOP 4 Consistent immersion depth See table on page 4 5 Pipette vertically or within 20 of vertical 6 Don t invert or lay the pipette flat with liquid in the tip Pre Rinsing Recommended Some soluti...
Page 8: ...ume delivered is actually 8 5 L 5 samples Change the volume setting to 11 8 L and repeat the measurements If the volumes deliv ered are still not close enough to 10 L make anoth er slight volume adjus...
Page 9: ...r arm snaps in place Pipet Lite Storage After use store the pipette in a clean safe place Pipet Lite is a precision instrument and should be treated with the level of care appropriate for laborato ry...
Page 10: ...a small amount of grease to the seal 4 If piston corrosion or staining is evident do not use the instrument Call 800 662 7027 for Service Leaks Inaccuracy Abnormal Stroke 1 Loose shaft Tighten couplin...
Page 11: ...is St Emeryville CA 94608 USA Tel 800 662 7027 Fax 510 652 4403 Rainin Service Center Rainin Road Woburn MA 01801 USA Tel 800 662 7027 Fax 781 935 7631 Rainin Europe Nieuwenhuizenweg 9 2314 XP Leiden...
Page 12: ...2 0 0 024 6 0 0 012 1 0 2 7 0 027 1 3 0 013 2 0 1 5 0 030 0 7 0 014 10 L 1 0 0 02 2 5 0 025 1 2 0 012 5 0 1 5 0 075 0 6 0 03 10 0 1 0 0 1 0 4 0 04 20 L 2 0 02 7 5 0 15 2 0 0 04 5 3 0 0 15 0 9 0 045 10...
Page 13: ...140 6200 145 6200 147 6200 157 6200 160 6200 169 C Short Shaft n a n a 6200 382 6200 383 n a D 6200 133 6200 133 6200 144 6200 148 6200 156 6200 163 6200 168 D Short Tip Ejector n a n a 6200 384 6200...
Page 14: ...B C D E F G H J A B C D E F G H J A B C D E F G H J Legend for Pipet Lite 2 L to 2000 L Models this page and top of next page A Plunger Button B Shaft Coupling C Shaft D Tip Ejector E Piston Assembly...
Page 15: ...G Pipet Lite 5000 L Pipet Lite 10 mL A C H H D E F G A C D E F G Legend for Pipet Lite 5000 L and 10 mL below A Plunger Button C Tip Ejector Arm D Piston Assembly E Piston O ring F Cylinder G Cylinder...
Page 16: ...nin com Pipette Service Phone 800 662 7027 Fax 781 935 7631 E mail service rainin com Direct Order Line Phone 800 472 4646 Fax 781 938 1152 E mail pipets rainin com Web www rainin com From outside the...