Rainin Instrument P-10 Manual Download Page 4

2

A stainless steel tip ejector is provided with all models except P-
5000 and P-10ML, for safe disposal of used tips. The ejector can
be removed by pulling on the plastic collar on the ejector arm.
Reattachment is quick and easy.

Autoclaving

The shaft and tip ejector are autoclavable: 121°C, 1 bar, 15–20
minutes. (P-5000, P-10ML do not use tip ejectors.)

Volume Indicator

The digital volume indicator is read from top to bottom. For P-2,
P-10, P-20, P-100, and P-200, black digits indicate microliters
and red digits tenths and hundredths of microliters. For P-1000
and P-5000, red digits indicate milliliters and black digits micro-
liters. For P-10ML, black digits indicate milliliters and red digit
tenths of a milliliter.

Sample values, volume ranges, and smallest increments for each

Pipetman model are shown below:

Pipetman

Range (µL)

Smallest

Model

Adjustable

Recommended

Increment µL

P-2

0 to 2

0.1 to 2

0.002

P-10

0 to 10

0.5 to 10

0.02

P-20

0 to 20

2 to 20

0.02

P-100

0 to 100

10 to 100

0.2

P-200

0 to 200

50 to 200

0.2

P-1000

0 to 1,000

100 to 1,000

2.0

P-5000

0 to 5,000

500 to 5,000

2.0

P-10ML

0 to 10,000

1 mL to 10 mL

20.0

Safety Filter, P-5000 and P-10ML

Models P-5000 and P-10ML use a safety filter in the end of the
shaft to help prevent liquid entering the shaft and contacting the
piston, should the plunger be allowed to snap up during aspira-
tion. This is particularly important using large volumes. If the filter
gets wet, replace it. 

For P-5000 insert the small diameter into the shaft; for 
P-10ML insert the large diameter into the shaft. 

P-20

12.5

µ

L

 1

 2

 5

P-100

75

µ

L

 0

 7

 5

P-200

125

µ

L

 1

 2

 5

 0

 7

 

 5

P-1000

0.75mL

7.5mL

 1

 2

 5

P-5000

1.25mL

P-10ML

P-10

7.5

µ

L

 0

 7

 5

P-2

1.25

µ

L

 1

 2

 5

 0

 7

 5

Red Digits

Black Digits

Summary of Contents for P-10

Page 1: ...Pipetman continuously adjustable digital microliter pipettes 8 pipettes for volume ranges from 0 1 L to 10 mL P 2 P 10 P 20 P 100 P 200 P 1000 P 5000 P 10ML Pipetman P 200 shown...

Page 2: ...and Repair 10 Contacting RAININ 10 Performance Specifications 11 Replacement Parts Back Cover FIGURES Figure 1 Pipetman P 200 1 Figure 2 Operating Pipetman 4 Figure 3 Removing Tip Ejector Arm 9 DIAGR...

Page 3: ...which reads in microliters The digital volume indicator simplifies volume setting and virtually elimi nates calculation errors All Pipetman models have a highly polished stainless steel piston patente...

Page 4: ...rements for each Pipetman model are shown below Pipetman Range L Smallest Model Adjustable Recommended Increment L P 2 0 to 2 0 1 to 2 0 002 P 10 0 to 10 0 5 to 10 0 02 P 20 0 to 20 2 to 20 0 02 P 100...

Page 5: ...he desired setting turn the dial 1 3 revolution higher than desired and reset the volume The friction ring prevents unintentional volume changes 3 Attach a new disposable tip to the pipette shaft Pres...

Page 6: ...p Selection I Tips must seal properly on the shaft to assure an air tight seal and avoid leaks or poor accuracy I Tips must be soft and flexible so that the shaft is not scratched or worn prematurely...

Page 7: ...mm P 200 50 200 L 3 6 mm P 1000 1000 1000 L 3 6 mm P 5000 500 5000 L 6 10 mm P 10ML 1 mL 10 mL 6 10 mm Tip immersion depth is important If exceeded the volume measured will be inaccurate possibly out...

Page 8: ...e any excess liquid from the outside of the tip without touching the orifice 5 To dispense rest the end of the tip against the vessel wall and press the plunger to the first stop Hold this position a...

Page 9: ...ntact with corrosive fumes may corrode the piston This will result in premature seal wear and may require refinishing or replacement of the piston Exposure of internal Pipetman com ponents to corrosiv...

Page 10: ...803D which use a grease seal Leaks Inaccurate Sampling 1 Loosened shaft Tighten coupling by hand 2 Split or cracked shaft Remove the tip ejector and inspect the shaft for fracture or split end Replace...

Page 11: ...ector opening over the end of the Pipetman shaft 5 Push the plastic collar into the handle until the tip ejec tor is securely in place Make sure that tips will fit prop erly on the shaft Figure 3 Tip...

Page 12: ...Tel 858 320 0446 It is NOT necessary to recalibrate the pipette after changing the seal or shaft Recalibration of the pipette is only necessary when the piston is replaced and should only be done by q...

Page 13: ...r both the density of water and evaporation where applicable Consult the Rainin publication Procedure for Evaluating Pipette Accuracy and Precision AB 15 for further information or download a copy fro...

Page 14: ...e diagrams are for dual adjustment Pipetman and cannot be used on older Pipetman models For convenience old style part numbers are also listed on the back page 12 P 2 P 10 P 20 P 100 A B M L K J H G F...

Page 15: ...13 P 200 P 1000 P 5000 P 10ML A B M L K J H G F C D E A B G F C D E A B D N N A B F G D G F...

Page 16: ...a n a n a n a 6190 164 6190 164 Filters 1000 n a n a n a n a n a n a 6190 165 6190 165 This column is for P 10ML after S N N05806D For P 10ML pre S N N05803D same as listed above except Plunger rod 23...

Reviews: