![Rae GammaRAE II Responder User Manual Download Page 12](http://html1.mh-extra.com/html/rae/gammarae-ii-responder/gammarae-ii-responder_user-manual_3369930012.webp)
0
GammaRAE II Responder
If the low gamma sensor is damaged or not functioning
properly, the message “Gamma sensor failed" is
displayed, turn the unit off and contact your distributor
or the RAE Systems service department.
The high gamma sensor (PIN diode) is also tested
continuously for the first 30 minutes of operation. If
the high gamma sensor is damaged or not functioning
properly, the message "HiGamma sensor failed"
is displayed, turn the unit off and contact your
distributor or the RAE Systems service department.
3. The datalog type is displayed, followed by the
alarm mode. These parameters can be changed in
Programming Mode, described later in this guide.
or