EchoView
User’s Guide
6
Marking
The EchoView is certified according to the IECEx scheme, ATEX
and CSA for US and Canada as protected by intrinsic safety.
The product is marked with the following information:
RAE SYSTEMS
3775 N. 1
st
. St., San Jose
CA 95134, USA
FMC-400
Serial No/barcode: XXX-XXXX-000
IECEx TSA
10.0016X
Ex ia IIC T4
Ex ia I
0575
IM1/II 1G
Ex ia IIC T4
Ex ia I
DNV 10 ATEX 83388X
Class I, Gr. A,B,C,D
T4
Intrinsically Safe
-40º C ≤ Tamb ≤ 50º C
Warnings:
Use only RAE 500-0111-000 for internal battery. Do not replace
internal battery in hazardous locations.
Ui: 3.6V; Ci: 78µF; Li/Ri: 3.5µH/ohm
Understand User’s Manual before operating or servicing.
Substitution of components may impair intrinsic safety.
To prevent ignition of flammable or combustible atmospheres,
disconnect power before servicing.