8
8. COMMUNICATION PROTOCOL
The display supports frame formats sent in a form of continuous
transmission.
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
I
space
Stability
marker
space
character
mass
space
unit
CR
LF
Stability marker
[space] if stable
[?] if unstable
[^] if error of exceeding range to + occurs
[v] if error of exceeding range to - occurs
Character
[space] for positive values
[-] for negative values
Mass
9 characters with decimal point and right justification
Unit
3 characters with left justification
Command
3 characters with left justification