If the standard is to be weighed in a container, the container needs
to be put on the pan and tarred,
•
Press (Estimate standard mass), then a message is displayed:
<Put standard>
,
•
Put the load on the pan. After stabilization the result is taken as
a standard (symbol
). Confirm it by pressing
,
•
At the same time the weight unit is changed to (w %).
25.3. Rederence unit mass inscribing into the memory
Procedure:
•
Enter
<
Deviations>
according to ch. 25.1 of this manual,
•
Press (Give piece mass), then an editing field is displayed
<Give piece mass>
together with the screen keyboard,
•
Enter a value and confirm by pressing
,
•
At the same time the weight unit is changed to (w %).
26. WORKING MODES – COMPARATOR
The
<
Comparator>
working mode allows users to eatimate standard
deviation for a series of measurements. The standard deviation is
calculated according the sequence of measurements ABBA or ABA:
A – control mass standard,
B – tested mass standard.
Number of measurements for one series and the method: ABBA or ABA,
are set by users in parameter group
<
Working modes>
in submenu
<
Comparator>
(see ch. 22).