![RADWAG WPF User Manual Download Page 34](http://html1.mh-extra.com/html/radwag/wpf/wpf_user-manual_781123034.webp)
Pr_n noStAb
- immediate printout
(not accessible in verified scales)
Pr_n StAb
- sending stable results
Pr_n rEPL
- automatic operation
Pr_n CntA
- continuous transmission in basic unit
Pr_n Cntb
- continuous transmission in present unit
Return to weighing:
see 10.2.2.
13.2. Minimal mass threshold
This function is necessary while working with
automatic tare
or
automatic operation or weighing animals
.
Automatic tarring
will not be applied until the indication (gross)
is lower than the value inscribed in
S_Lo
parameter.
In automatic operation
measurements (net) are sent via RS232
when the indication is equal or greater than the value inscribed
in
S_Lo
parameter.
Weighings animals
is performer when the indication is equal or
greater than the value inscribed in
S_Lo
parameter.
Procedure:
•
Enter the submenu
<P2.Prnt>
and then:
- 34 -
Summary of Contents for WPF
Page 2: ...2 JANUARY 2012...
Page 11: ...WPT H1series pillar 7cm main dimensions WPT H2 WPT H6 series main dimensionS 11...
Page 13: ...WPT C1 R series main dimensions WPT C1 series main dimensions 13...
Page 14: ...WPT C2 K series main dimensions WPT C2 R series main dimensions 14...
Page 15: ...WPT C2 series main dimensions WPT C3 K series main dimensions 15...
Page 35: ...Return to weighing see 10 2 2 13 3 Baud rate Procedure Enter the submenu P2 Prnt and then 35...