7
Software of counting set makes possible:
-
determining mass of single detail on main balance with highest
accuracy of weighing
-
owing to application of graphic display, balance software is user-
friendly
-
possibility of application PC type PS/2 keyboard simplifies moving in
software menu and data accessing
-
printing basic labels, cumulative labels, „cc labels” (cumulative labels of
cumulative ones)
-
cooperation with barcode scanner
-
cooperation with additional balance platform
3. GETTING STARTED
3.1 Conditions of appropriate use
9
place balance on a stable and flat table or free from vibrations
9
balance should not be exposed to draughts and sudden air movements
9
balance should be placed in room of stable temperature and humidity
9
balance should be placed far from heat sources
9
temperature of the room should be +15
°
C
÷
+30
°
C (for WLT balances)
9
if static electricity influence balances, indications, its base should be
earthed. Earthling clamp is in the back part of balance base
9
balances should be levelled according to a level condition indicator to
guarantee the appropriate weighing accuracy
3.2 Time of warming-up
Before performing measurements wait until balance reaches temperature
stabilization. It is so-called warm-up time. It takes about 15 minutes.
For weighing instruments that have been stored in low temperatures (e.g.
during winter) the warm-up time is about 2 hours.
The indications of display may change during temperature stabilization.