33

 

19.  WORKING MODES – General Information 

 

The weighing device features the following working modes:  

 Weighing, 

 Parts 

counting, 

 +/- 

control, 

  Percent weighing %, 

 Peak 

Hold, 

 Totalizing, 

 Animal 

weighing. 

 
 

19.1. Running Working Mode 

  Go to home screen, press 

 key, name of the first available working 

mode is displayed. 

 Press 

 

key, names of available working modes are displayed 

successively one by one. 

  Enter selected working mode, to do it press 

 key. 

 

 

The weighing device program has been designed to make 
the scale run, upon restart, with the latest operated working 
mode on. 

 
 

19.2. Working Modes Local Settings 

 

Each working mode features specific (local) functions which enable adapting 
device operation to individual needs. The functions are to be found in local 
settings. To go to local settings of each working mode enter 

<P3.Func>

 

submenu. Some special functions are available for all working modes, refer  
to the table below: 

 

 

Accessibility 

Save mode 

Time interval 

Lo threshold 

Weighing 3.1.1.Acc 

3.1.2.Snn 

3.1.3.Int 

3.1.4.Lo 

Parts counting 

3.2.1.Acc 

3.2.3.Snn 

3.2.4.Int 

3.2.5.Lo 

+/- control 

3.3.1.Acc 

3.3.2.Snn 

3.3.3.Int 

3.3.4.Lo 

Percent weighing % 

3.4.1.Acc 

3.4.3.Snn 

3.4.4.Int 

3.4.5.Lo 

Peak Hold 

3.5.1.Acc 

3.5.2.Lo 

Totalizing 3.6.1.Acc 

3.6.2.Snn 

3.6.3.Int 

3.6.4.Lo 

Animal weighing  

3.7.1.Acc 

3.7.3.Lo 

Summary of Contents for WLC/C2

Page 1: ...WLC USER MANUAL ITKU 103 04 11 18 EN WLC C2 Precision Scale WLC F1 Precision Scale...

Page 2: ...2 NOVEMBER 2018...

Page 3: ...0 2 Operating the Menu 15 10 3 Return to the Weighing Mode 16 11 WEIGHING OPERATION 16 11 1 Good Weighing Practice 16 11 2 Zeroing 17 11 3 Taring 17 11 4 Entering Tare Value Manually 18 11 5 Dual Rang...

Page 4: ...OUNTING 36 21 1 Local Settings 36 21 1 1 Selecting Operation Mode 36 21 2 Setting Sample Mass by Entering Mass of a Single Part 36 21 3 Setting Sample Mass by Determining Mass of a Single Part 37 22 W...

Page 5: ...n A Prior the first use carefully read this User Manual Use the device only as intended B Indicators to be decommissioned must be decommissioned in accordance with valid legal regulations 2 2 Battery...

Page 6: ...tion to polarization Close the lid 3 WARRANTY CONDITIONS A RADWAG feels obliged to repair or exchange all elements that appear to be faulty by production or by construction B Defining defects of uncle...

Page 7: ...ried out outside RADWAG authorized service point service claims intrusion into mechanical or electronic construction by unauthorized people the device does not bear company security stickers F Warrant...

Page 8: ...8 Dimensions of WLC F1 R scale Dimensions of WLC C2 K scale Dimensions of WLC C2 R scale...

Page 9: ...4 3 Pins Overview Pin2 RxD Pin3 TxD Pin4 5VDC Pin5 GND RS232 1 connector DB9 M male Pin2 RxD Pin3 TxD Pin4 5VDC Pin5 GND RS232 2 connector DB9 M male 5 UNPACKING AND INSTALLATION Unpack the device an...

Page 10: ...2 Connecting the Scale to the Mains The weighing device can be connected to the mains only with a power supply that comes standard with the particular model Nominal voltage of the power supply specifi...

Page 11: ...n may cause indication errors To remove the indication errors perform user adjustment 6 4 Battery Charge Status The scale of standard design is equipped with an internal battery Battery state is signa...

Page 12: ...aning Glass Components Select dissolvent depending on a dirt Never soak the glass panes in alkaline solutions since they interact with glass and may cause damage Do not use cleansers containing abrasi...

Page 13: ...d a little bit of dishwashing detergent 7 4 Cleaning Powder Coated Components For preliminary cleaning process stage you need running water or wet sponge featuring large holes this will help you to re...

Page 14: ...tion key press to change the working mode Press to send the weighing result to a printer or computer Press to zero the scale Press to tare the scale Upon pressing keys combination functions of given k...

Page 15: ...n Communication P5 ducE Peripheral devices P6 Prnt Printouts P7 Othr Operation related functions P8 InFo Scale data P9 Unit Units 10 2 Operating the Menu In order to navigate the program menu use the...

Page 16: ...o the home screen press key repeatedly 11 WEIGHING OPERATION Load the weighing pan Read the result when stability marker is displayed Only stable weighing results can be recorded stability marker 11 1...

Page 17: ...g To determine net weight value load the weighing pan with a packaging wait for a stable indication and press key Zero indication and the following pictograms are displayed Net and The weighing device...

Page 18: ...I weighing range Upon switching to weighing with the accuracy of the II weighing range respective pictogram marker is displayed in the top left hand corner To return to weighing with accuracy of the...

Page 19: ...it disabled for verified weighing devices Select start unit and press key next go back to the home screen to do it press key Upon next start up the scale runs with set start unit 11 6 2 Temporary Unit...

Page 20: ...carried out prior weighing if long breaks between successive measuring series occur if the ambient temperature has changed dynamically if the scale has been relocated Adjustment types external adjust...

Page 21: ...ht mass is displayed the mass value must be 30 of the maximum capacity value Enter weight mass value and press key to confirm text UnLoAd remove weight is displayed Remove the load from the weighing p...

Page 22: ...onfirm next go to the home screen The higher filter value the longer the weighing takes 13 2 Value Release Enter this parameter to adjust rate of stabilisation of the measurement result Depending on t...

Page 23: ...nded to disable the function Procedure Enter P2 rEAd 2 4 Aut submenu Press key parameter values are displayed successively one by one YES autozero function enabled no autozero function disabled Press...

Page 24: ...lect to enter tare value manually by means of keys combination tArnn Select to enter tare value that is stored in scale memory use keys combination Press key to confirm next go to the home screen 13 7...

Page 25: ...Net symbol is shown in the upper left corner of the screen The tare value acquired from the weighing device memory is not remembered upon the weighing device restart 13 8 Last Digit Function designed...

Page 26: ...odd Press key to confirm next go to the home screen 14 2 RS232 2 Port Settings Enter P4 Conn 4 2 rS2 submenu and set respective transmission parameters 4 2 1 bAd Baud rate 2400 4800 9600 19200 38400...

Page 27: ...sion in basic unit Cntb Continuous transmission in current temporary unit Press key to confirm next go to the home screen 15 1 3 Printout Interval for Continuous Transmission Parameter enabling you to...

Page 28: ...arameter values are displayed successively one by one nonE none rS1 RS232 1 rS2 RS232 2 Press key to confirm next go to the home screen 16 PRINTOUTS It is possible to define adjustment report printout...

Page 29: ...t data and signature fields 6 1 7 SiG An area for the signature of an operator carrying out the adjustment Printouts are generated exclusively in English Report example 16 2 GLP Printout P6 2 GLP is a...

Page 30: ...English Report example 17 MISCELLANEOUS PARAMETERS P7 Othr is a group of parameters enabling you to adapt the scale to individual needs 17 1 Backlight Parameter allowing you to change display brightn...

Page 31: ...device shuts down automatically If the indication is stable during the declared time interval the device is shut down automatically Shutdown function is inactive and the device cannot be turned off i...

Page 32: ...H 24 hour format 17 5 Default Operator Settings Parameter allowing you to restore default operator settings Procedure Enter P7 Othr 7 8 dFLu submenu text Cont is displayed Continue Press key to confir...

Page 33: ...rking mode on 19 2 Working Modes Local Settings Each working mode features specific local functions which enable adapting device operation to individual needs The functions are to be found in local se...

Page 34: ...weighing device to a peripheral device Procedure Enter P3 Func menu and select given working mode Go to Snn function Press key parameter values are displayed successively one by one StAb Manual print...

Page 35: ...he mass indication must get below the set net value of Lo threshold Procedure Enter P3 Func menu and select given working mode Enter Lo function window for entering Lo threshold value is displayed Ent...

Page 36: ...o Lo threshold For detailed description read section 19 2 4 21 1 1 Selecting Operation Mode Parameter allowing you to select method of determination of sample piece mass Procedure Enter 3 2 PcS 3 2 2...

Page 37: ...ss key to confirm first text LoAd is displayed for 1 second then the weighing window If the parts are to be weighed in a container first put the container on a weighing pan and tare it Load the weighi...

Page 38: ...trol first text SEt Lo is displayed for 1 s next window for declaring low weighing threshold Min Enter respective value and press key to confirm first text SEt Hi is displayed for 1 second next window...

Page 39: ...3 3 4 5 Lo Lo threshold For detailed description read section 19 2 4 23 1 1 Selecting Operation Mode Parameter allowing you to select method of determination of reference weight mass Procedure Enter 3...

Page 40: ...ing pan during one weighing process 24 1 Local Settings To go to local settings enter 3 5 toP submenu 3 5 1 Acc Working mode accessibility For detailed description read section 19 2 1 3 5 2 Lo Lo thre...

Page 41: ...ising Operation Enter Add working mode Totalising blinking pictogram is displayed If the ingredients are to be weighed in a container first put the container on a weighing pan and tare it Load the wei...

Page 42: ...weighing pan 26 WORKING MODE ANIMAL WEIGHING Animal Weighing is a working mode enabling you to weigh products that disrupt efficient establishing of stability It is mostly intended to measure weight o...

Page 43: ...an animal On exceeding the set mass value of Lo threshold parameter animal weighing starts this is signalled with display of dash Upon process completion mass value of an animal is snapped and display...

Page 44: ...stalled or installed incorrectly Check if batteries are installed correctly polarization The scale switches off automatically 7 3 t1 parameter set to value enforcing scale shut down after particular t...

Page 45: ...value of single part greater than maximum capacity in Parts counting working mode Entered value of Max threshold greater than maximum capacity in control mode Entered reference mass greater than maxim...

Page 46: ...46...

Reviews: