45
•
Set respective value and press
key to confirm, next go to the home
screen.
23.2. Reference Sample Mass Determined by Weighing
•
Enter
<3.4.dEu / 3.4.2.UUt>
submenu, set
<
S_S
>
value.
•
Enter
<dEu>
working mode (Percent weighing), first, text
<LoAd>
is displayed for 1 second, then the weighing window.
•
Load the weighing pan with the reference sample. When the indication
is stable (
pictogram is displayed), press
key to confirm the mass.
•
Mass of the weighed load is automatically set as reference sample mass,
the home screen is displayed along with
100.000%
value.
23.3. Reference Sample Mass Determined by Entering the Mass Value
•
Enter
<3.4.dEu / 3.4.2.UUt>
submenu, set
<Suu>
value.
•
Enter
<dEu>
working mode (Percent weighing).
•
Text
<SEt_Ut>
is displayed for 1 s, next, window for declaring mass
of reference sample.
•
Enter respective value and press
key to confirm. The home screen
is displayed automatically with
0.000%
value.
If the value of entered reference sample mass is greater than
max capacity value, then message <Err Hi> is displayed.
24. WORKING MODE – PEAK HOLD
Peak Hold is a working mode allowing you to snap value of maximum force
applied to the weighing pan during one weighing process.
24.1. Local Settings
To go to local settings enter
<3.5.toP>
submenu.
3.5.1.Acc
Working mode accessibility
For detailed description read section 19.2.1.
3.5.2.Lo
Lo threshold
For detailed description read section 19.2.4.