Thus prepared anti-draft chamber and panes can be properly cleaned. All the operations have to be
done carefully. Pay special attention to the spot where the weighing pan is installed: dirt and other
small elements might enter the balance construction through this opening, which might negatively
influence the balance parameters.
Upon completion of the cleaning process assembly the balance, take actions like before but in a
reverse order. Pay a special attention to the left and the right side pane, assembly it on the correct
side of the balance. While sliding the pane back onto its place remember to pull the protection (like
before when dismantling the anti-draft chamber).
3.5. POWERING THE DEVICE
The balance has to be plugged to the mains using the original power adapter exclusively. Nominal
power supply of the power adapter (specified on the power adapter data plate) has to be compatible
to the power from the mains. Plug the power adapter’s to the power supply socket located at the back
of the housing.
3.6. TEMPERATURE STABILIZATION
Before measurement it is necessary to wait until the balance reaches thermal stabilisation.
For balances that were stored in much lower temperatures before plugging to the mains, thermal
stabilisation period shall take at least 12 hours. During the thermal stabilization, the indications on a
display panel can change. It is recommended that ambient temperature changes at place of use were
insignificant (slow to change).
3.7. CONNECTING PERIPHERAL EQUIPMENT
Use only accessories and peripheral equipment recommended by the manufacturer. The balance
must be unplugged before connecting or disconnecting any peripherals (printer, PC computer,
computer keyboard). On connecting the peripherals, plug the balance to the mains.
3.8. BALANCES FEATURING TERMINAL-WEIGHING MODULE WIRELESS CONNECTION
(an option)
The terminal operates on batteries, wireless connection ensures its cooperation with the weighing
module. This, in contrast to a standard design, allows to eliminate cable connecting the devices.
Wireless connection solution is marked by letter
<B>
, e.g.
XA 220.4Y.B
.
Features:
Maximum range: 10 m
Maximum operation time of the terminal (without the necessity to charge batteries): 8 h
Advantages:
1.
Lack of additional sources of vibration
Terminal located elsewhere than the weighing table eliminates vibrations being a result of
operation. This ensures stable working conditions for balances with the highest resolution. No
specific weighing workstations are needed.
2.
Comfort of weighing in fume cupboards and laminar flow cabinets