62
Readout
For detailed description read section 20.1.
Save mode
For detailed description read section 20.2.
Auto Threshold
For detailed description read section 20.3.
ACAI
For detailed description read section 22.2.1.
Minimum Reference Sample Mass
For detailed description read section 22.2.2.
Result control
For detailed description read section 20.4.
Autotare
For detailed description read section 20.5.
Labelling mode
For detailed description read section 20.6.
Statistics
For detailed description read section 20.7.
22.2.1. Function of Automatic Correction of Reference Sample Mass
<ACAI>
is a specific function enabling you to correct weight value of a single
piece by means of the weighing device program.
Procedure:
•
Enter
<Working modes / Parts counting / ACAI>
submenu and set
respective option.
Where:
Function of automatic correction of reference sample mass disabled.
Function of automatic correction of reference sample mass enabled.
<ACAI>
function gets activated for
<Parts counting>
mode at the moment
of determination of reference sample quantity, it is signalled by display
of
pictogram at the home screen top bar.
There are four conditions of ACAI function operation implemented into
the weighing device program:
1. the weighing result must be stable,
2. pieces quantity must increase,
3. pieces quantity upon adding cannot increase more than twice,
4. pieces quantity upon adding must be comprised within
±
0.3 tolerance
of total value.
If you decide that the reference sample features the right amount of pieces, you
can record weight value of a single piece to weighing device memory and
deactivate the function by pressing
key.