Procedure:
After displaying the level of batteries (in per cents) the program returns
to weighing.
14.4.2. Battery discharge pictogram
The symbol (bat low) switches on when the voltage level drops to 18%
of the accepted level of voltage. It means that charging or exchanging
batteries is required.
Low level of batteries:
•
Pictogram
on the display,
•
After one time the device will automatically switch off to protect
the batteries from distructable discharging,
•
Charging is signalled by
(blinking period about 2 seconds)
on the display.
14.4.3. Accumulator charging option
This function allows to switch on charging algorithm for
NiMH
batteries (for indicators plastic casing):
a)
Parameter
<CHr6>
set to
<no>
:
•
Pictogram
does not appear, charging disabled,
•
During software initializing, after turning on
„bAtt”
.
b)
Parameter
<CHr6>
set to
<YES>
:
•
Pictogram
blinks slowly (period about 2 seconds),
charging is enabled,