- 31 -
P2.2.3 – PRINT MODE
P2.2.3.1 – MODE [EACH / WHEN STAB. / AUTO]
P2.2.3.2 – AUTO. PROG
P2.2.4 – INFORMATION [TARE / NETTO / GROSS / USER / PRODUCT / /NSTD.
INF. / NONE]
P2.2.5 – INF. NSTD. (non-standard information – any text, max 19 characters)
P2.2.6 – HOT KEYS (available options depending on the working mode)
P2.2.6 1 – F1 [NONE / ENTER TARE/ …]
P2.2.6.2 – F2 [NONE / ENTER TARE / …]
P2.2.6.3 – F3 [NONE / ENTER TARE/ …]
P2.2.6.4 – F4 [NONE / ENTER TARE / …]
9.4.1 Filter level setting
Filter shall be set depending on the ambient and environment conditions. For the
best possible conditions the filter can work in a very fast mode (V.FAST value for
parameter P.2.2.1.1 Filter); however, if the conditions are bad (shakes, drafts), the
filter should be set to slow or very slow option (SLOW or V. SLOW value for
parameter P.2.2.1.1 Filter). The effectiveness of the filter is different for the
weighing range. The filter works slower when “approaching” the weighed mass, it
works more intensively for weighed mass within the set range of the filter
(parameter for setting filter range is accessible only from the service menu – the
user does not have an access to it).
Depending on the filter, the weighing time is shorter (V.FAST and FAST) or longer
(SLOW and V. SLOW).
CAUTION!
The higher filter level, the longer weighing time.