- 68 -
18. ERROR MESSAGES
Er1 Hi
- wrong initial mass or calibration
Er2 nuLL
- values from A.D converter out of the normal range - under range
Er3 FuL1
- values from A.D converter out of the normal range - over range
Er4 FuL2
- over range
Er5 rout
- value out of range for performed function e.g value > +/- 2% from initial
mass for zeroing, e.g. value <= 0 for tarring
Er7 tout
- timeout error for zeroing and tarring
Er8 outr
- enter value out of range
Er9 Lock
- function locked e.g. to protect function if LFt (legal for trade) is on,
Er10 cal
- calibration error (change of mass or incorrect mass of standard mass).
19. ADDITIONAL EQUIPMENT
19.1. Vibration damping table
Thanks to this stable base all vibrations are eliminated.Inside of
table marble plate is installed which makes foundation for balance.
19.2. Additional display
Characteristics: Length of conductor between additional display
and balance – 1,5m, plastic casing, possibility of bending the
display.
19.3. Computer software
PW-WIN 2004 – gathering data from any balance made by
RADWAG.
RAD-KEY 2000 – enables to introduce result of weighing to Exel.
REC-FS 2000 – making mixtures. It has its own programmed
contents base according to recipes are made.
Number of instruction:
LMI-26-02/06/08/ENG