Cleaning the analyzer
24
The analyzer must always be kept clean. Exterior surfaces, the inlet gasket and other
parts of the analyzer must be cleaned when they are contaminated with blood and/or
other liquids.
Note:
Radiometer has not tested whether cleaning wet wipes can be used for this.
Prerequisite: A lint-free cloth.
Further instructions can be found in the latest version of "ABL90 FLEX Instructions for use"
(under "Replacements and maintenance").
Cleaning the inlet gasket
The inlet gasket must be kept clean to make sure that samples can be aspirated
correctly.
WARNING – Risk of infection
Make sure you do not prick or scratch yourself on the inlet probe.
• Stress: the inlet probe must not be bent – replace it if it is
To clean the analyzer exterior
Prerequisite: A lint-free cloth.
1. Lightly dampen a lint-free cloth with water.
2. Put your finger on a part of the screen that is not active and hold it there.
3. Gently wipe the screen.
Note:
The sensor cassette compartment and the top surface of the solution pack
compartment must be cleaned by a Radiometer service representative.
Disinfecting
Follow your local, state and federal guidelines. For further instructions, please refer to the
latest version of “ABL90 FLEX Instructions for use” (under “Replacement and maintenance”).