Standard unit
incompatible
Manual reprocessing error:
The calibration method used for the
reprocessing has a standard definition different from the calibration
curve stored.
Edit the parameters of the calibration method to be reprocessed
(Standard menu).
Start addition
Define when the reagent addition will be performed.
Start addition = Initial addition
The addition is performed at the start of the titration (titrant predose
included).
Start addition = After EP no. x (or IP no. x)
The addition is performed after EP no. x (or IP no. x). The titration to
the next EP (or IP) starts at the end of the time entered for Delay after
addition.
Enter in:
Edit method > Parameters (if Addition = Yes)
Edit reagent > Calibration parameters (if Addition = Yes)
Initial addition is the only option available in Continuous IP
mode.
Start from
Command of the Sequence/Sample stack window that enables to
start a TIM or SAC sequence from a beaker number of your choice.
Refer to "Sequence/Sample stack menu", page 229
.
Refer to "Running a TIM sequence", page 218
.
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...