Molar weight
When the result unit = weight (g, mg, µg, g/l, mg/l, g/kg, mg/kg,%),
enter the molar weight in g/mol of the species to be determined.
Enter in:
Edit method > Results
Range available:
0.001 to 1000 g/mol.
Monotonic IP
method
For this type of method, a preset volume of titrant (increment) is
added to the solution as soon as the measurement variations satisfy a
stability criterion or at the end of a preset acceptation time.
Figure 29: Monotonic IP method - size of the increments
In Monotonic IP, one measurement is saved per increment added.
Select a Monotonic IP method when you want to minimise the number
of increments added and in some conditions to speed up a sample
titration.
The Monotonic IP method is stopped after a given number (1 to 8) of
equivalent points has been detected.
Nat. water
(temperature
correction)
Refer to "Temp. correction None/Linear/Nat. water", page 246
.
E or pH
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...