TitraLab 870 Reference Manual
Page 135
GRAN =
f(Volume) curve
Definition:
Refer to "GRAN representation:", page 271
.
Displaying the GRAN = f(volume) curve:
At the end of an ISE standard addition test, the instrument displays
the concentration and the slope calculated at 25°C:
Printing the GRAN = f(volume) curve:
The curve is printed automatically at the end of each test if asked for
in the Printouts menu of the ISE standard addition method,
see "Printouts setup", page 179
Example of printout:
Press
2
More details then
4
Curve.
The GRAN = f(Volume) curve is
displayed. The sample and addition
points are marked on the curve (here 3
additions have been performed).